We've detected that you're using an ad content blocking browser plug-in or feature. Ads provide a critical source of revenue to the continued operation of Silicon Investor.  We ask that you disable ad blocking while on Silicon Investor in the best interests of our community.  If you are not using an ad blocker but are still receiving this message, make sure your browser's tracking protection is set to the 'standard' level.
Biotech / Medical : Chromatics Color Sciences International. Inc; CCSI
CCSI 28.05+5.2%Oct 30 3:59 PM EDT

 Public ReplyPrvt ReplyMark as Last ReadFilePrevious 10Next 10PreviousNext  
To: Gurupup who wrote (4557)9/9/1998 10:49:00 AM
From: JanyBlueEyes  Read Replies (4) of 5736
 
Pediatrics 1998 Sep;102(3):E28

Noninvasive estimation of serum bilirubin.

Tayaba R, Gribetz D, Gribetz I, Holzman IR

Department of Pediatrics, City Hospital Center at Elmhurst, Queens, New York.

[Medline record in process]

Objective. The aim of this study was to evaluate the accuracy and clinical usefulness of a new computer-driven, hand-held device (Chromatics Colormate III) to estimate serum bilirubin from skin-reflectance (skin color) of neonates.

Study Design. A total of 2441 infants (both term and premature) at two hospitals had repeated measurements of transcutaneous bilirubin. Of these infants, 900 had one or more laboratory determinations of serum bilirubin. Many of the infants had early measurements of skin color before the onset of jaundice.

A visual estimate of the degree of jaundice was made by a health care worker when the laboratory study was drawn. A subgroup of 61 infants was also studied while undergoing phototherapy with a total of 284
comparisons obtained.

The reproducibility of the instrument was assessed separately using
standardized color tiles and repeated measurements by multiple operators.

Results. The range of serum bilirubin measurements that had concurrent skin color measurements was 3.22 to 338.1 &mgr;mol/L (0.2 to 21 mg/dL).

The linear regression indicated an r = 0.956, and 95% of the skin color measurements were within 32.2 &mgr;mol/L (2.0 mg/dL).

There was no interference with the accuracy of the device because of infant race or weight, or because of the use of phototherapy.

The device provided reproducible information when infants were tested repeatedly over 30 minutes; the coefficient of variation for the transcutaneous bilirubin measurement was 3.1% around a mean estimate of 135.32 &mgr;mol/L (8.4 mg/dL).

Conclusion. The Chromatic Colormate III allows for a clinically useful estimate of serum bilirubin in a wide variety of infants. By using a color discrimination algorithm and obtaining a skin measurement before the onset of icterus, this instrument can provide valuable clinical information that obviates the need for serum bilirubin determinations. Its use in newborn nurseries may allow physicians to shorten length of stay more safely and decrease the use of invasive blood tests.

PMID: 9724676, UI: 98393764