We've detected that you're using an ad content blocking browser plug-in or feature. Ads provide a critical source of revenue to the continued operation of Silicon Investor.  We ask that you disable ad blocking while on Silicon Investor in the best interests of our community.  If you are not using an ad blocker but are still receiving this message, make sure your browser's tracking protection is set to the 'standard' level.
Technology Stocks : Helix Technology, a cold play on semiconductor equipment
HELX 38.08+0.3%Nov 26 4:00 PM EST

 Public ReplyPrvt ReplyMark as Last ReadFilePrevious 10Next 10PreviousNext  
To: John Finley who wrote (833)4/28/2000 8:45:00 AM
From: j t jackson   of 1227
 
JF... Re: GM cooler on GE MRI systems. Coolers have been used for some time now (15? years) to eliminate liquid nitrogen boil-off and to provide an additional heat sink in the 10 degree Kelvin area to intercept structural heat leaks, heat leaks in the electrical leads and as an additional radiation shield to minimize the liquid helium boil-off rate. If the magnet operates at 4.2K, the boiling point of liquid helium, then you would have to make a trade off between an oversized refrigerator for a reasonable cool-down time of the large magnet mass or a long cool-down time for a refrigerator sized to maintain the operating thermal load. Would be interesting to know the magnet operating temperature. Regards..jtj